Universität Erlangen-Nürnberg, Institut für Organische Chemie
Processing of JEOL 2D-NMR spectra with ACD software
FT of HETCOR spectra
- Open file XXX_VCHSHF.gxd
- click "FT along t2"
- Choose these options:
- complex
- make zero-filling, e.g. from initial 1024 to final 2048
- apodization function: "user", click "EM", set LB=2
- => ok
- click "done" (left top edge)
- click "FT along t1"
- important:"Dataset type" must be set to "AV complex reversed"
(Note: this "mirrors" the spectrum along f1; you may also employ the
"Process => reverse spectrum => along f1" tool after complex FT)
- apply "linear prediction", "forward"
- points to predict e.g. 129 to 256; base points 1 - 128, coeff. count 8
- apodization function "Pseudo Echo"
- => ok
- click "done" (left top edge)
Don't hesitate to play around with processing parameters, e.g. linear prediction
instead of zero-filling in t2; compare spectra with those obtained by routine output
FT of COSY spectra
- Open file XXX_VCOSYNH.gxd
- click "FT along t2"
- Choose these options:
- complex
- make zero-filling, e.g. from initial 1024 to final 2048
- apodization function: "sine bell"
- => ok
- click "done" (left top edge)
- click "FT along t1"
- "Dataset type": default
- apply "linear prediction", "forward"
- points to predict e.g. 129 to 256; base points 1 - 128, coeff. count 8
- apodization function "Sine Bell"
- => ok
- click "done" (left top edge)
Don't hesitate to play around with processing parameters, e.g. linear prediction
instead of zero-filling in t2; compare spectra with those obtained by routine output
How to plot high-resolution 1D spectra at a 2D spectrum
- Important: 1D spectrum must be processed and stored before it can be plotted
to a 2D spectrum!
- select "Tools =>setup 1D curves"
- click "Horz" in top bar
- select appropriate 1D spectrum (XXX_BCM.ESP = 13C; XXX_NON.ESP = 1H)
- adjustment: move cursor into 2D-spectrum
- press "Strg"
- press left mouse button, drag left => 1D-spectrum will move right
- fine-adjust 1D-spectrum to correct cross peaks, use zoom function if needed
- repeat adjustment along f1:
- click "Vert"
- click "Spectrum"
- select XXX_NON.ESP
- press "Strg"
- move cursor into 2D plot
- press left mouse button, drag mouse cursor down => spectrum will move up
- click green "ok" (top left edge) when adjustments are finished
NMR Department
Institute of Organic Chemistry
Location: Rooms U34, U35, U36
University of Erlangen-Nürnberg
Henkestrasse 42
D-91054 Erlangen, Germany
Phone +49-(0)-9131-85-22987
FAX +49-(0)-9131-85-22991
e-mail:
bauer (at) chemie.uni-erlangen.de